Semi-Automatic
Blood Pressure Monitor
NOW with Date & Time Model 1103
|
||||
|
|
||||
|
Distributed by:
The Lumiscope Co., Inc
Repair Department: 1-800-672-8293
Made and Printed In China
|
User Manual
|
|||
|
|
||||
|
|
||
![]() |
||
|
|
||
|
|
|||
|
Table of Contents
|
|||
|
|
|||
|
Lumiscope guarantees this product free from defects in material and workmanship for a period of two years from the date of purchase, except as noted below:
This Lumiscope product warranty does not cover damage caused by misuse or abuse, the attachment of any unauthorized accessory, alteration to the product, or any other conditions whatsoever that are beyond the control of Lumiscope. Lumiscope shall not be responsible for any type of incidental, consequential, or special damage. All implied warranties, including, but not limited to those implied warranties of fitness and merchantability, are limited to the total duration of two years from the original date of purchase.
To obtain warranty service on your Lumiscope Blood Pressure Monitor, please contact the Lumiscope Repair Department at 1-800-672-8293 or visit www.lumiscope.net . There is a $15 fee for return shipping and handling. Please make checks payable to The Lumiscope Co., Inc. Upon receipt, Lumiscope will repair or replace, as appropriate, this blood pressure monitor and return it to you. Warranty is solely through the Lumiscope Repair Department. Service of this product by anyone other than the Lumiscope Repair Department voids warranty.
In the event that Lumiscope does not abide by the terms of this warranty, the consumer may seek breach of contract remedies in the New York State Federal courts of law. This warranty gives you specific legal rights and you may also have other rights which vary from state to state.
Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
|
Introduction.....................................................................3
About Blood Pressure................................................4, 5
Parts Identification..........................................................6
Display Symbols & Explanations....................................7
Installing the Batteries....................................................8
Correct Measuring Posture............................................9
Applying the Cuff.................................................... 10, 11
Helpful Tips................................................................... 12
Measurement Procedures......................................13, 14
Memory Recall............................................................. 15
Deleting the Memory.................................................... 16
Clock Settings.............................................................. 17
Care & Maintenance.................................................... 18
Troubleshooting............................................................ 19
Specifications............................................................... 20
Warranty....................................................................... 21
|
||
|
|
|||
|
|
||||
|
Introduction
|
||||
|
|
||||
|
Congratulations on your purchase of the new Lumiscope Model 1103. This semi-automatic blood pressure monitor is both easy-to-use and ideally suited for daily measurements. The large display includes systolic blood pressure reading, diastolic blood pressure reading and the pulse rate, all clearly shown at the completion of each reading.
You can store up to 85 measurements, ideal for users who want to monitor and track their blood pressure on a regular basis. The 1103 is compact and portable, making it perfect for home use and travel.
Blood pressure measurements determined with your 1103 are equivalent to those obtained by a trained observer using cuff/stethoscope auscultation method, within the limits prescribed by the American National Standard, Electronic or Automated Sphygmomanometers.
Please read this manual carefully before use. For specific information on your own blood pressure, contact your physician. Please be sure to keep this manual.
|
Model
Display System ON/OFF source
|
1103
Liquid Crystal Display
4 "AA" Alkaline batteries
|
||
|
Measurement Method: Oscillometric
Measurement Range : Pressure: 20~300 mmHg
Pulse: 40-200 beats/minute
|
||||
|
Accuracy
Inflation
Pressure Exhaust Memory
|
Pressure: ±3 mmHg
Pulse: within ±5% of reading
Inflation Bulb
Manual release valve
85 sets
|
|||
|
Low Battery Indicator : Yes
|
||||
|
Auto shut-off
Battery Life
Operating Environment
Storage Environment
Weight
Arm circumference Dimensions
|
3 minutes after last key operation
Around 2000 operations
10°C~40°C(50°F~104°F) < 85% RH.
-10°C~50°C (-4°F~122°F) < 85% RH
Approximately 259g (.57 lbs) with batteries
8.5"-14"
5.5" (L) x 3.9" (W) x 2.9"(H)
|
|||
|
|
||||
|
*Specifications are subject to change without notice.
|
||||
|
|
||||
|
|
||||||||||||||||||||||||||||||||||||||||||
|
About Blood Pressure
|
||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||
|
What is Blood Pressure?
Blood pressure is the pressure exerted on the artery tube while blood flows through the arteries. The pressure measured when the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the heart dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure.
Why Measure Your Blood Pressure?
Among the various health problems afflicting people today, problems associated with high blood pressure are by far the most common. High blood pressure's dangerously strong correlation with cardiovascular disease has made measuring blood pressure a necessity for identifying those at risk.
|
|||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
![]() |
About Blood Pressure
|
![]() |
Care & Maintenance
|
|||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
Blood Pressure Standard
The World Health Organization (WHO) and National High Blood Pressure Education Program Coordinating Committee have developed a blood pressure standard, according to which areas of low and high-risk blood pressure are identified. This standard, however, is a general guideline and blood pressures vary between different people, age groups, etc.
It is important that you consult with your physician regularly. Your physician will tell you your normal blood pressure range as well as the point at which you will be considered at risk.
|
|
|||||||||||||||||||||||||||||||||||||
|
Grade 3 Hypertension (Severe)
|
||||||||||||||||||||||||||||||||||||||
|
160-
|
Grade 2 Hypertension (Moderate)
|
|||||||||||||||||||||||||||||||||||||
|
I
|
40-
|
Grade 1 Hypertension (Mild)
|
||||||||||||||||||||||||||||||||||||
|
High-Normal
|
||||||||||||||||||||||||||||||||||||||
|
120-
|
Normal
|
|||||||||||||||||||||||||||||||||||||
|
Optimal
|
110
|
|||||||||||||||||||||||||||||||||||||
|
85 90 100
Diastolic Pressure mmHg
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||
|
Parts Identification
|
||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||
|
Setting the Clock, Date & Time
|
||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||
|
A. Press and hold the "MEMORY" button for over 3 seconds. The display will show a blinking year. Press the "ON/ OFF" button to adjust the year when it's blinking.
|
\ /
ED 05-
/ \
|
![]() |
||||||||||||||||||||||||
|
||||||||||||||||||||||||||
|
B. Press the "MEMORY" button again. The "MONTH" will appear and blink to be adjusted, the "DATE", "HOUR" and "MINUTE" will follow as you press and release the "MEMORY" button.
|
||||||||||||||||||||||||||
|
||||||||||||||||||||||||||
|
C. Press the "ON/OFF" button to adjust the values of the "MONTH", "DATE", "HOUR" and "MINUTE" when they are blinking.
|
![]() |
|||||||||||||||||||||||||
|
||||||||||||||||||||||||||
![]() |
||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
|
![]() |
Deleting the Memory
|
![]() |
|||||||||||||||||||||||||||||||||||
|
|
Display Symbols & Explanations
|
|||||||||||||||||||||||||||||||||||||
|
I
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
Display Symbols
|
To delete the memory
1. To delete one memory:
A. Press the "MEMORY" button to display the record that you wish to delete.
B. Press and hold the "ON/OFF" button until the display shows "DEL".
|
![]() |
||||||||||||||||||||||||||||||||||||
|
Q 3^ memory QO_/^_
irur on
|
Systolic Pressure
|
|||||||||||||||||||||||||||||||||||||
|
no
|
IB
|
|||||||||||||||||||||||||||||||||||||
|
Diastolic Pressure
|
||||||||||||||||||||||||||||||||||||||
|
dEL
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
Pulse Rate
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
Display Explanations
|
C. Once the "DEL" has been displayed, press and hold the "ON/OFF" button again. After 3 beeps, the selected record has been deleted.
2. To delete all memories: Follow steps A and B above
|
|||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||
|
C. Press the "MEMORY" button until the display shows "DEL ALL".
|
![]() |
|||||||||||||||||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||||||||||||||||
|
D. Once the "DEL ALL" has been displayed, press and hold the "ON/OFF" button again. After 3 beeps, all records have been deleted.
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||
![]() |
Memory Recall
|
![]() |
Installing the Batteries
|
![]() |
|||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||
|
To see the previously measured values
A. Press the "MEMORY" button to recall the last reading. The LCD display will show the memory number, actual date and time simultaneously of the measurement.
|
To insert the batteries
A. Press down on the latch and lift the cover towards you to open the battery compartment.
B. Insert 4 "AA" batteries as indicated. Make sure to match polarities (+, -) on the batteries.
C. Replace the battery cover by inserting the two hooks at the bottom into the slot, then push on the top area.
|
||||||||||||||||||||||||||||||||
|
Memory Number ^
|
![]() |
|
|||||||||||||||||||||||||||||||
|
Note: Please remove batteries when not in use for an extended period of time. Do not dispose batteries with household garbage.
|
|||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||
|
Date ->
|
![]() |
![]() |
|
||||||||||||||||||||||||||||||
|
Time ^
|
Q MEMORY
s: a l?Q BE IB
|
||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||
|
B. To view additional readings, push the "MEMORY" button successively to go through the various readings. (1~85).
|
|||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
Correct Measuring Posture
|
![]() |
Measurement Procedure
|
![]() |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
A. Place your elbow on a table so that the cuff is at the same level as your heart.
Note: The level of your heart is slightly below your armpit.
|
![]() |
C. After the cuff pressure reaches the recommended pumping value, allow the unit to release gradually while performing the measurement.
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Relax your entire body, especially
the area between your elbow and fingers.
|
Fig. A
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Pumping
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
B. If the cuff is not at the same level as your heart or if you can not keep your arm completely still throughout the reading, use a soft object such as a folded towel to support your arm.
C. Turn your palm upwards.
D. Sit upright in a chair, and take 5-6 deep breaths.
Note: Avoid leaning back or crossing your legs while the measurement is being taken. (Fig C)
|
![]() |
Note: It is important to remain still and quiet during measurement. Any significant movement may affect measurement results.
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Results
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
D. Once the measurements have been displayed, press the air release valve to automatically deflate the cuff. Then press "ON/OFF" button to turn off the unit or it will automatically shut off after 150 seconds.
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Fig. B
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
![]() |
Air Release Valve
|
![]() |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
ON/OFF
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Fig. C
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
![]() |
Measurement Procedures
|
![]() |
Applying the Cuff
|
![]() |
||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
To start the measurement
Note: In order to get your correct measurement result and, to avoid erroneous measurements, you should be seated quietly. Your arm, hand, and fingers should be still during the measurement procedure.
A. You must depress the knob on the inflation bulb to release the air inside the cuff before each measurement. Press the "ON/OFF" button. The display will initially show a flashing "0".
|
To apply the cuff
A. Plug the end of the tubing on the cuff into the connector at the base of the monitor.
B. Unwrap the arm cuff leaving the "Index" end of the cuff through the D-ring of the cuff.
|
|||||||||||||||||||||||||||||||||||||
![]() |
![]() |
|||||||||||||||||||||||||||||||||||||
![]() |
![]() |
— Flashing "0'
|
||||||||||||||||||||||||||||||||||||
|
Initial Display
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
B. After the "0" stops blinking, start to pump the inflation bulb several times until the reading reaches the recommended pumping value. ( See below).
|
C. Remove any clothing that restricts your upper left arm. Wrap the cuff around your upper left arm. The rubber tube should point in the direction of your hand. Position the artery mark over the main artery (on the inside of your arm) in the upper arm.
|
|||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||
![]() |
||||||||||||||||||||||||||||||||||||||
|
Main Artery
|
||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|
|
|||
|
D. Turn your left palm upward and place the edge of the arm cuff at approximately 1-2 inches above the inner side of the elbow joint. Tighten the cuff by pulling the end of the cuff.
|
Here are a few helpful tips to help you obtain a more accurate reading.
A. Do not measure your blood pressure immediately after consuming a large meal. To obtain more accurate readings, please wait one hour before measuring.
B. Do not smoke or drink alcohol before measuring your blood pressure.
C. You should not be physically tired or exhausted while taking a measurement.
D. It is important that you relax during measurement. Take a 15-minute rest before a reading.
E. Do not take measurements if you are under stress or under tension.
F. Take your blood pressure at normal body temperature. If you are feeling cold or hot, wait a while before taking a measurement.
G. If the monitor is stored at very low temperature (near freezing), place it at a warm location for at least one hour before using it.
H. Wait about 5 minutes before taking the next pressure measurement.
|
||
![]() |
|||
|
E. If the Index line falls within the OK range indicated at the edge of the cuff, this cuff is suitable for your use. If the Index line falls out of the OK range, you may need a special cuff for measurement. Please contact our service department at 1-800-672-8293 for more information regarding additional cuffs.
|
|||
![]() |
|||
|
Note: To locate the main artery, press two fingers approximately 1-2 inches above the bend of your elbow on the inside of your left arm.
|
|||
|
|
|||